What are the solutions for large temperature fluctuations in high and low temperature test chambers?
Author:
Fault phenomenon:The low temperature of the high and low temperature test chamber is unstable and fluctuates greatly.
Fault analysis:
First,observe whether the refrigeration compressor of the high and low temperature chamber can start during the operation of the high and low temperature test chamber.If the compressor of the high and low temperature chamber can start during operation,it indicates that the electrical circuits from the main power supply to the compressors of each high and low temperature chamber are normal,and there is no problem with the electrical system.When there are no issues with the electrical systems of the high and low temperature chamber,continue to inspect the refrigeration system.First,inspect the two sets of refrigeration compressor units in the high and low temperature chamber.If the exhaust and suction pressures of the low-temperature(R23)stage compressor are both lower than the normal values,and the suction pressure is in a vacuum state,it indicates that the refrigerant charge of the main refrigeration unit is significantly insufficient.Touch the exhaust and suction pipelines of the R23 compressor in the main unit of the high and low temperature chamber with your hand again.

If you find that the temperature of the exhaust pipeline is not high and the temperature of the suction pipeline is not low(no frost has formed),this also indicates that the R23 refrigerant in the main unit is lacking.One is the main unit and the other is the auxiliary unit of the high and low temperature test chamber.When the cooling rate is relatively high,both units work simultaneously.At the beginning of the temperature maintenance stage,both units still work simultaneously.When the temperature initially stabilizes,the auxiliary unit stops working,and the main unit maintains the cooling and stability of the temperature.If the primary unit R23 leak,can make the main units of cooling effect is not big,because the cooling process,the two units work at the same time,so no temperature stable and not like now,cooling rate,reduce and instructions.During the temperature maintenance stage,once the auxiliary unit stops working and the main unit has no cooling effect,the air temperature inside the high and low temperature test chamber will rise slowly.When the temperature reaches a certain level,the control system will start the auxiliary unit again to cool it down,and then the auxiliary unit will stop working.This process repeats,resulting in the fault phenomenon that the low temperature cannot be maintained.
At this point,it has been confirmed that the cause of the failure of the high and low temperature chamber is the leakage of the refrigerant R23 in the low-temperature(R23)stage unit of the main refrigeration unit.High and low temperature test chamber maintenance methods:Check for leaks in the refrigeration system of this high and low temperature test chamber.Use a combination of a leak detector and soapy water to identify the location of the leak.If it is found that the valve stem of the hot gas bypass solenoid valve is cracked with a fine gap,replace this solenoid valve.If leaks are found in other areas,use oxygen welding to complete the welding of the leakage points,and then refill the system with fluorine.The system operation can then return to normal.

After comprehensive analysis,for the fault phenomenon that the low temperature of the high and low temperature test chamber cannot be maintained,the analysis and judgment are basically carried out in a way from easy to difficult,from the outside to the inside,and from electrical to air conditioning.Familiarity and understanding of the principle and working process of the high and low temperature test chamber are the basis for analyzing and judging faults.Only by thoroughly understanding the working principle and process of the high and low temperature test chamber can various problems that occur during its operation be quickly solved.