Common faults and handling methods of constant temperature and humidity test chambers.

Author:

GUANGDONG SHUANGYI TESTING EQUIPMENT CO., LTD.


The constant temperature and humidity test chamber,also known as the constant temperature and humidity testing machine,constant temperature and humidity test box,programmable wet heat alternating test chamber,constant temperature machine or constant temperature and humidity box,is a device used to test the performance of materials under various environmental conditions and to test the heat resistance,cold resistance,dry resistance and moisture resistance of various materials.It is suitable for quality inspection of electronic,electrical,mobile phone,communication,instrument,vehicle,plastic products,metal,food,chemical,building materials,medical,aerospace and other products.

Due to the fact that constant temperature and humidity test chambers are generally heavy,large in size,inconvenient to move or relocate,and have complex circuits,when faults occur,it often makes it difficult for customers to handle them.In order to enable each customer to quickly identify and solve faults by themselves in a timely manner,the following introduces some solutions for the three major faults in the maintenance of constant temperature and humidity chambers.

I.The temperature and humidity display of the constant temperature and humidity test chamber is abnormal

1.The humidity display of the temperature and humidity controller in the constant temperature and humidity chamber is abnormal.At this time,we should first check the over-temperature protector that controls the humidifier to see if it is still at the 130℃position,and ensure that the abnormal humidification is not caused by the water temperature protection setting being too low.

2.The second examination indoor line is too long didn't clean,causing drainage unobstructed,not again to check whether there is water,wet cloth tank inlet pipe road is unobstructed,line presence of air.

3.Solid state relay have final inspection humidity is punctured,the cause has been humidification.We can use a clamp-on ammeter to check whether the humidifier is working properly and whether its power is normal,etc.

4.Shut off the main power of constant temperature and humidity box,the humidifying barrels below the location of the drain valve open to parallel to the line,will be the inside of the water discharge,clean the internal dirt,and then pull out the silicone pipe,rub in the water,after the maintenance according to dismantle the way,back in the opposite direction.After the installation is completed,the drain valve at the bottom of the humidification bucket should be closed to a position perpendicular to the pipeline,and then the main power supply at the rear of the machine should be turned on to conduct a humidity test.

II.How to solve the overpressure alarm

First of all,we should confirm the ambient temperature,because the vast majority of overpressures are caused by excessively high ambient temperatures.Confirm whether the room temperature is too high due to summer or poor ventilation.Generally,the suitable temperature range for a constant temperature and humidity test chamber is 5 to 35 degrees Celsius.

If the room temperature does not exceed what has been mentioned above,when placing the constant temperature and humidity test chamber,is the equipment placed too close to the wall or the corner?Because this will lead to poor heat dissipation and cause an alarm.

Another factor to consider is whether the test chamber has been cleaned since it was purchased.Generally,beneath the machine,that is,in the refrigeration system,one can see large refrigeration component accessories such as the condensing fan and condenser.Check if the condensing fan and condenser are already covered with dust.If so,please clean it up in time,because these tiny and unobtrusive dust particles are preventing it from dissipating heat.It is recommended to clean it once a month or once a quarter on a regular basis.When cleaning,you can use a vacuum cleaner or other auxiliary tools.Please note that the cleaning must be carried out when the machine is off.

Inspection:The reasons for poor heat dissipation of the condenser are:

(1)The condenser is too dirty.

(2)With air cooling system.

(3)The surrounding air flow.

(4)The ambient temperature is too high.

(5)The condenser inner surface oil film.

The processing result:The cooling effect of this test chamber is normal at night,which rules out reasons(1),(2),(3)and(5)mentioned above.If the test chamber is moved to another place,it will no longer"trip".It can be determined from this that the poor heat dissipation of the condenser is caused by the constant temperature and humidity test chamber being close to the heat source.

Solution:Move the constant temperature and humidity test chamber to a place far from heat sources and well-ventilated,and the fault can be eliminated.

III.The wet heat system malfunctions

Possible manifestations include:the humidity in the constant temperature and humidity chamber shows 100%,and the humidity cannot be reduced when the low humidity point is set.What should be done when encountering these situations?Humidity rises slowly or fundamental rise will not come up the solution of specific as follows:

1.Check whether the humidity OUT on the controller has an output or a 100%output.

2.See if SSR lights light up,if you have light,you can use clip-on table measure current.If there is no current,it could be due to the following reasons:

A.The circuit breaker tripped,cutting off the heating path.

B.Humidification,solid state relay SSR is damaged,has signal input but can't conduction;

C.Humidification pipe damage.

If there is an input for the humidity signal on the controller but the signal indicator light on the SSR does not light up,it is necessary to focus on checking the water level detection float ball of the humidification water pan(or humidification boiler).Check whether the float ball does not float due to insufficient water level or is damaged when the water level is sufficient,which leads to the humidification signal not being delivered to the SSR and the humidification cannot be carried out.

Low humidity slow down or not drop the solution of the specific as follows:

1.Check the compressor work,if there is a job,check if high voltage is lower than 6 bar,high and low pressure table may have leakage fluorine case,need to add fluoride treatment.

2.Check studio for air blow out,if there is no or air volume is small,may be a motor reversal,the power phase sequence fault.The power cord needs to be commutated.

If have other question,welcome to contact Mrs.Huang for details,we will try our best to help you.