The Shuangyi testing equipment in the field of environmental reliability testing, accurately simulating various temperature and humidity conditions is of vital importance. The Double-Layer Constant Temperature Humidity Test Chamber, also known as the double-door constant temperature and humidity test chamber or the upper and lower layer constant temperature and humidity chamber, is precisely an efficient and innovative solution, specially designed to meet the high requirements of modern laboratories for testing efficiency and space utilization.
Key purchasing considerations:
1. Temperature range and accuracy: Clearly define the required high and low temperature limits as well as the control accuracy requirements.
2. Humidity range and accuracy: determining the maximum/minimum humidity and the control precision.
3. Double control mode: independent control (can be run completely different application) or cooperative control (usually run the same or related program).
4. Within the box size: according to the sample size and the number under test, determine the lower level of their effective volume.
5. The uniformity of the temperature and humidity and volatility: core performance indicators, directly affects the reliability of test results (refer to the GB/T10586, GB/T 10592 standard).
6. Refrigeration method: air-cooled (convenient) or water (high efficiency, need water).
7. Controller function: steps program, cycle times, communication interface (USB, Ethernet and RS485), data recording ability, etc.
8. Safety protection: perfect over temperature, overload, water shortage, leakage, etc. Multiple protection mechanism.
Technical parameters of the Shuangyi double-layer constant temperature and humidity test chambers
|
Model |
YPH-64L |
YPH-80L |
YPH-100L |
YPH-150L |
YPH-225L |
YPH-408L |
YPH-800L |
YPH-1000L |
|
Inner size |
W40*H40*D40cm |
W40*H50*D40cm |
W50*H50*D40cm |
W50*H60*D50cm |
W60*H75*D50cm |
W80*H85*D60cm |
W100*H100*D80cm |
W100*H100*D100cm |
|
Outer size |
W65*H140*D110cm |
W65*H155*D110cm |
W75*H155*D135cm |
W75*H165*D145cm |
W85*H180*D145cm |
W105*H190D155cm |
W125*H215*D175cm |
W125*H215*D195cm |
|
Control system |
Balanced temperature and humidity control system |
|||||||
|
Temperature range |
-70℃~+150℃(180℃)(Full non-linear & no-load operation) -60℃~+150℃;-40℃~+150℃;-20℃~+150℃;0℃~+150℃(Full non-linear & no-load operation) |
|||||||
|
Humidity range |
20%~98%RH |
|||||||
|
Temperature fluctuation |
±0.5℃ |
|||||||
|
Humidity fluctuation |
±2.5%RH |
|||||||
|
Temperature deviation |
±2.0℃ |
|||||||
|
Humidity deviation |
±3%RH (≥75%RH);±5%RH (<75%RH) |
|||||||
|
Heating time |
-20℃~+150℃about 45min; -40℃~+150℃about 60min;-70℃~+150℃about 75min |
|||||||
|
Cooling time |
20℃~-20℃about 40min; 20℃~-40℃about 60min; 20℃~-70℃about 85min |
|||||||
Core concept: the efficiency of the space and the dual optimization of the structure of "Double layers" is the biggest bright spot. This design is not merely a simple superposition of top and bottom, but rather refers to the fact that the equipment has two independent or separable test Spaces (upper and lower layers) inside, sharing a set (or independent yet highly efficient and collaborative) precise temperature and humidity control system. This is equivalent to achieving the functions of two independent test chambers within the floor area of one device.
Double-layer structure implemented:
1. Independent circulation system: each chamber has an independent fan, air duct and/return air outlet, ensure their uniformity of the temperature and humidity, each other. The control system can independently set and run different temperature and humidity programs for the upper and lower layers.
2. Efficient coordination system: Shared core cooling/heating humidifying module, but through the precision air valve and the zone control, by individual or synchronization to run different operating mode, the lower level optimization efficiency.
3. The physical separation: Double-Layer Constant Temperature Humidity Test Chamber through the solid insulation layer separated into two independent test chamber.
4. Air duct design: the unique is the key to optimize air duct system.
The remarkable advantages:
1. Saves valuable space: One unit occupies the space of two single-box devices, achieving the functions of two separate devices, significantly optimizing the layout of the laboratory space.
2.Enhance test efficiency: Two sets of the same or different temperature and humidity tests can be conducted simultaneously, significantly improving equipment utilization and test throughput.
3.Reduces purchase cost: Compared to purchasing two independent single-box devices, it usually has a higher cost-effectiveness.
4. Reduces energy consumption: Sharing the core temperature and humidity generation system (especially the collaborative control type), it can lower the overall operating energy consumption.
5. Facilitates operation and management: Centralized control and management of two sets of tests, reducing the workload of operators.
Typical application scenario:
1. Electronic electrician: Testing different batches of PCB, components at the same time, the machine under high and low temperature, damp and hot on the performance and reliability.
2. Auto parts: Concurrent validation a variety of interior trim materials, electronic control unit (ECU), sensors of tolerance in extreme temperature humidity environment.
3. Materials research: a comparative study of two kinds of new materials (such as plastic, coating, composite) under the condition of the same or different aging performance change.
4. Bio-pharmaceutical: Can be medicines and chemical reagents, reagent or synchronous packaging materials under different conditions of temperature and humidity storage stability test (must be in accordance with the relevant specification).
5. Scientific research institutes: Efficient support multiple research projects or student experiment demand for temperature and humidity environment.
6. Quality inspection institutions: to improve the batch sample detection ability, meet the demand of customer diversified test.
The China Shuangyi double-layer constant temperature humidity test chamber, through innovative spatial layout and advanced control technology, effectively resolves the contradiction between the tight laboratory space and the improvement of testing efficiency while ensuring the testing accuracy of a single chamber. It provides an efficient and reliable environmental simulation testing platform for multiple industries such as electronic products, materials research, automobiles, and bio-pharmaceuticals, and is an ideal choice for enhancing R&D and quality inspection capabilities.
To explore your lab double layers constant temperature humidity test chambers’ solutions? Our technical team is always ready to provide you with professional consultation and jointly formulate equipment solutions that meet your unique testing needs.

Keywords
Further details


Team Highlights
Guangzhou Jiaoguang Asia Exhibition
Guangzhou Jiaoguang Asia Exhibition
Shanghai Munich Exhibition
Company Qualifications
Double-Layer Constant Temperature Humidity Test Chamber
Contact Information
Category