Some common troubleshooting tips for thermal shock test chambers
Author:
1.In the high-temperature test of the thermal shock test chamber,if the high temperature does not reach the test temperature value:check the electrical system and eliminate faults one by one.If the temperature rises slowly,check the air circulation system to see if the regulating baffle is opening normally.If it cannot be opened,then check whether the motor of the air circulation is operating normally.
2.During the heating process,there is a severe temperature overshoot:readjust the setting parameters of the PID.If the temperature rises directly,over-temperature protection will occur.At this point,it may be that the controller has malfunctioned and the control instrument needs to be replaced.
3.If the thermal shock test chamber suddenly malfunctions during operation:Check the fault prompt on the controller,quickly identify whether there is a fault at the relevant position according to the prompt,and report the fault issue to the manufacturer.Handle it under the guidance of the manufacturer.If it cannot be handled,ask the manufacturer to arrange for maintenance.
4.The low temperature of the thermal shock test chamber cannot be reached:If the cooling is slow,it may be that too many samples are placed in the test area,and the cold air cannot circulate fully.If the temperature shows a tendency to rise after reaching a certain value,this is due to the poor usage environment of the equipment.The ambient temperature where the equipment is placed and its location(the distance between the back of the box and the wall)must meet the requirements.
5.The target humidity of the thermal shock test chamber varies greatly:it might be caused by the drying of the gauze on the wet bulb sensor.First,check if the water level controller is normal.If it is,just use the gauze.
6.If the humidity is much lower than the set value,mainly check whether the humidification system is working,examine the water supply system of the humidification system to see if there is a certain amount of water in the water supply system,and whether the water level control of the humidification boiler is normal.Secondly,check the electrical control system to see if there is any abnormality in the control system that causes the water supply system to fail to work normally.